ROSEMOUNT 400A PARTS LIST (SIMILAR TO BECKMAN 400) (HYDROCARBON ANALYZER)

Part Number Description
25B0001R0 400: Power supply cover
25B0002R0 400: Temperature Controller
400A-2008-0001 400A, 120 VAC on alarm bus
898816 400A: range relay {*}
623983 400A: Range Select Board
9385K61 400A: Foam Insulation
25C0008R0 400A: Display driver
400A ANALYZER 400A: Rosemount Analyzer
25A0001R0 400: Glow plug / igniter
25C0001R0 400A light up procedure
25C0007R0 400A: 4-20 mADC board
400A-ALARM 400A: alarm module
25C0005R0 400A: cable 16 conductor
25C0009R0 400A: Door, Inside plexiglass
25C0004R0 400A: high range capillary
25C0003R0 400A: mixed fuel capillary
25C0002R0 400A: pure fuel capillary
620443RU 400A: RANGE SWITCH BD
400A-REPAIR 400A: repair of analyzer
620423-REPAIR 400A: Repair of Preamp Board
73659 400: Burner Tip Assembly
630375 400: Cap assy w/glow plug
73658 400: collector assy
73674 400: Electrode kit
617900 400: Glow plug / igniter (*)
834341 400: Lamp, Pilot {*}
643459 400: SS GAUGE, 5psig
6A00324G01 400A: Main Elect Bd Updated
656888 400A: Air restrictor blue dot
620428 400A: BOARD, AMPLIFIER *
621031 400A: CABLE, 16 COND
655178 400A: Door,Inside plexiglas{*}
620446 400A: Flame Out Bd (Obsolete)
656921 400A: high range capillary {*}
860918 400A: igniter toggle switch{*}
630669 400A: Mixed fuel restrict red
620423 400A: Preamp Board
620443 400A: RANGE SWITCH BD
898904 400A: reed relay {*}
904731 400A: reed relay 5V, 1A {*}
898861 400A: relay, 24v {*}
876447 400A: SPAN POT, 30K {*}
655171

25C0007R0 4-20 MADC ISOLATED OUTPUT BOARD (REPLACES OEM 620433 & 621023)

Description: This is an isolated signal analog output card used industry wide. The analyzer’s potentiometric output (DC volts) signal drives the signal for the 4-20 mADC card (usually 0VDC input = 4 mADC output & 5 VDC input = 20 mADC output). The term “isolated” refers to the fact that the output signal is not referenced to earth ground and is therefore much less susceptible to interference from other voltage sources. Milliamp signals are very useful when running long signal loops (on the order of tens of feet to thousands of feet long; note that total loop resistance has to be less than 850 ohms).

Where used: 400A,755,755A,755R,870

Static sensitive: This board is not very susceptible to ESD but handle it with some ESD protocols.

Replaces: Rosemount 620433 (in kit p/n: 621023)

Compatibility: RIGAS has released its compatible version of the OEM board. Some nice enhancements were included: [1] hook-type test points, [2] better labeling, [3] heat sinks, [4] zero & span potentiometers can dual configuration (i.e., adjustable from the top or from the front).

Please specify potentiometer type when ordering. Choices: [1] end adjust or [2] top adjust

Typical failure modes (after many years of service): [1] total failure: most likely due to the fullwave bridge rectifier or output opamp. [2] non-linearity: most likely due to the charge pump circuit. [3] inability to calibrate: most likely due to the charge pump circuit. [4] noisy output: output opamp is failing –or– input signal is noisy –or– board mounted isolated power supply(ies) are failing –or– loop resistance is very high –or– it could be that your mADC reader (DCS, DAS, etc) is misbehaving.

Repairable: Yes!  Call us at 877-616-0600 for an RMA number.