RIGAS – TELEDYNE LIGHTHAWK 560: QUARTERLY AUDIT PROCEDURE

Quarterly audits should be simple … thorough … but still as simple as possible. Unlike the Teledyne procedure (see below), our method is fully compliant with EPA Method 203, EPA Procedure 3, ASTM D 6216-98, and 40 CFR part 60.

This is a snippet of RIGAS’ procedure:

1. record the serial numbers
2. challenge the PLCF
3. note the Fault indicators
4. record current readings (including Dust Compensation)
5. record current calibration values
6. record calibration set points
7. record LED drive current
8. check purge air system (replace air filters as necessary [be careful not to disturb the dust, that will alter your EPA data])
9. align
10. perform a dust accumulation test (which cleans the optical surfaces as a PM measure)
11. manual calibration cycle check
12. record new Dust Compensation value
13. install Cal Kit Fixture
14. do NOT perform Background Set
15. do NOT perform Normal Set
16. do NOT remove Cal Kit Fixture
17. do NOT perform Zero Set
18. check Cal Zero value
19. do NOT manual calibration cycle check
20. analog output check
21. perform Calibration Error Test
22. record all EPA data from the data historian (or chart recorder)
23. clean the unit (outside)
24. complete the EPA compliant report

Please contact us if you’re looking for a vendor to do quality EPA audits, analyzer preventive maintenance, analyzer repair (on site or in our depot),parts, or telephone technical support. We also have Teledyne LightHawks to rent.

Leave a Reply